Calibration; The Calibration File; Creating A New Calibration File - Teledyne FALCON GC Operation And Maintenance Manual

Table of Contents

Advertisement

Falcon GC
1. With the Falcon touch screen
From the home screen press the green Start button. If the button
is not green then there are still some modules that have yet to
reach their setpoints. Check to make sure that any FID flames are
lit.
2. With the Chromperfect software:
From the Run window, under the Status tab, right-click
anywhere in the white Excel-like fields and select start.
The Falcon is now in operation mode.

4.2 Calibration

4.2.1 The Calibration File

A calibration file is shipped with your GC. However, there may be
instances where you need a new calibration file or you want to update
your current calibration file.

4.2.2 Creating a New Calibration File

First, make sure that you have a chromatogram that you would like
to use to create your calibration file. If you have a two-channel
instrument, these steps are the same for both channels and you should
create two separate calibration files.
Click the "C" Chromperfect icon to open the Chromperfect banner
and then click on "Analysis".
Teledyne Analytical Instruments
Operation
19