Adjusting Fid Pressures - Teledyne FALCON GC Operation And Maintenance Manual

Table of Contents

Advertisement

Falcon GC
if the flame is lit by holding a reflective surface over
the FID exhaust and examining for condensation.
Check with a flow meter to ensure that the Air (ignite)
Pressure, the Air (Run) Pressure and the Hydrogen
Pressure values are consistent with those found in your
Data Package shipped with the instrument.

5.2 Adjusting FID Pressures

To adjust the pressure of in the FID please follow the steps described
below.
Note: For Ambient and N
below may be different. Refer to the Tech Note provided
with the specific instrument.
1. Turn off the H
the flow to 5psi.
2. Click "Remote" on the Chromperfect screen.
3. Select the "Pressures" tab.
Teledyne Analytical Instruments
Maintenance & Troubleshooting
Carrier Gas applications, the settings
2
and air flow to the FID. If carrier gas is N
2
adjust
2
25