A bed for all circumstances that unifies technology and comfort. Technically perfected
down to the last detail: The practico alu plus and the practico alu SCC can be extended
without any tools, and therefore individually adapted to patients. It is really easy: The fully
ergonomic, real bed extension is inserted in the blink of an eye, as described at the end of
this chapter. The side rails are provided for all lengths. These nursing care beds provide a
high degree of comfort to frail people, patients who need care and people with disabilities,
thus facilitating an optimal care through a high lying comfort and an easy operation.
-
practico alu plus and practico alu SCC are not intended to be used in hospitals.
-
If the practico alu plus or practico alu SCC are equipped with open wheels, the
nursing care bed is suitable for the transport of patients. The bed is mobile with
the patient resting in it. To prepare for this, lock the castors and move the lying
surface to the lowest horizontal position. Unlock the castors and move the bed.
After the transport, lock the castors. If the nursing care bed is equipped with a
roller cover, the bed is only suitable to be moved for cleaning purposes inside the
patient's room or to allow access to the patient.
-
The nursing care bed is suitable for persons (adults) in need of care who are at
least 146 cm tall. The person's weight must not exceed 185 kg and must be over
40 kg. The body mass index (BMI) must be greater than or equal to 17.
-
Under certain circumstances the practico alu plus or practico alu SCC can be used
(if necessary) for medical purposes with other electric medical equipment (e.g.
suction devices, ultrasonic humidifiers, food systems, anti-bedsore systems, oxy-
gen concentrators and similar devices). In this event, disable all bed functions for
the duration of the application via the integrated disabling function.
Attention: The bed has no special connection options for a potential equalisation. Electri-
cal medical devices connected to the patient intravascular or intracardiac may not be
used. The operator of the medical products has to ensure that the combination of the
equipment meets the requirements of EN 60601-1.