※ This calibration function is for resetting incubator
temp. or humidity with user's thermometer or hygrometer.
As incubator is already calibrated from factory with
international standard, so it is not recommendable
to the user's calibrate by themselves. In case calibration
is required, users can do this with their thermometer or
hygrometer.
Spread egg tray mat on the bottom of tray,and stable
the incubator temperature by operating for an hour
before calibration.
And then, observe the temperature or humidity at
several position above 10mm of the hatching pad with
your thermometer, and calibrate with average value.
Low priced thermometers or hygrometers which you can easily purchase from the market have
large deflection of measurement. This could cause hatching failure, and it is recommended
to use highly precise thermometer or hygrometer.
Temperature Calibration
▶ Press
Then "CA" will indicated for about 0.5sec, and
it changes to current temperature.
(CA means Calibration.)
If you want to increase the temperature 0.5°C,
Please reduce 0.5°C in calibration setting mode.
▶ You can control temperature by 0.1°C with
and
▶ Press
it saves the calibrated value.
After that, it goes to humidity calibration.
If you use this function wrong, initialize
the incubator to factory setting.
DIGITAL INCUBATOR
Incubator Hatcher Rcom King suro MX20 User Manual
6. Product Information
,
and
button at the same time.
button.
button after calibration finished then
6. Product Information
(1) How to Calibrate
the Temperature & Humidity
Thermometer
or Hygrometer
Hatching Pad
Sensor
Egg Tray
10mm
Temp. UP
Temp. Down
(1) How to Calibrate
the Temperature & Humidity
25