Dimensions In Mm (Inch); Connection Diagram - E+E Elektronik EE211 Operation Manual

Heated humidity / temperature sensor
Hide thumbs Also See for EE211:
Table of Contents

Advertisement

Outstanding long term stability under high humidity conditions
The operation principle of EE211 copes with the causes for poor long-term stability of non-heated
sensors at continuously high humidity. The constant over-temperature of the EE211 sensing head
(approx. 5 °C = 9 °F) means max. 76 % RH humidity at the sensors and enables following benefits:
• The sensing head of EE211 stays dry even under condensing conditions, which prevents dust and dirt
from sticking to the sensor and leads to outstanding long-term stability.
• The combination of dry sensing head, E+E proprietary coating of the sensing element and sealed
solder pads minimize the impact of corrosive agents.
• Maximum humidity of 76 % RH at the sensor eliminates the drift caused by exposure to continuous
high humidity.
3.2

Dimensions in mm (inch)

Basis Unit:
90
± 0,3
"
+0.11
(3.54
)
5
(0.2")
Fig. 1
Dimensions of basis unit and temperature probe
3.3

Connection Diagram

EE211-M1J3 - Digital Output
Address switches
EE211-M1A2/3/5/6 - Analogue Output
Setup interface
Fig. 2
Connections and setup
6
Ø 13
(Ø 0.51")
Ø 12
(Ø 0.47")
Bus termination resistor 120 Ω (jumper)
Selection output signal U / I
Temperature Probe:
8
(0.3")
M12x1
15...35 V DC
24 V AC ± 20%
Output:
Modbus RTU
15...35 V DC
24 V AC ± 20%
Output: 0...5 V
0...10 V
0...20 mA
4...20 mA
Operating manual EE211 Heated humidity and temperature sensor
75
(2.9")
50
(2")
OUT1
V
OUT2
V
mA
mA