The user manual may not be used for the purposes of competition without the written consent of E+E Elektronik Ges.m.b.H. and may not be forwarded to third parties. Copies may be made for internal purposes. All information, technical data and diagrams included in these instructions are based on the information available at the time of writing.
Failing to follow the instructions in this user manual may lead to measurement inaccuracy and device failures. ▪ The EE212 may only be operated under the conditions described in this user manual and within the specification included in chapter 7 Technical Data. ▪...
Environmental Aspects PLEASE NOTE Products from E+E Elektronik Ges.m.b.H. are developed and manufactured in compliance with all relevant environmental protection requirements. Please observe local regulations for the disposal of the device. For disposal, the individual components of the device must be separated according to local recycling regulations.
EE212 sensor (hot-swap). Refer to chapter 6.1 Sensing Module Replacement for details. For a setup deviating from default, the EE212 can be configured manually by means of the free EE-PCS Product Configuration Software. Refer to chapter 5 Setup and Configuration for details.
Page 7
User Manual EE212 Type T13 Mounting flange in the scope of supply for type T2 Ø > 16 Ø > 0.63 0.24 Radiation shield HA010501 for type T13 (needs to be ordered separately) Ø105 (Ø4.13) 44 (1.73) 44 (1.73) 1.5 (0.06) 122 (4.8)
It is important to make sure that the cable glands are closed tightly for the power supply and outputs cable. This is necessary for assuring the IP rating of the enclosure according to EE212 specification, as well as for stress relief at the screw terminals on the EE212 board.
The display is available as spare part, intended for replacement only, not for retrofitting an EE212 originally ordered without display. PLEASE NOTE The EE212 (4 - 20 mA, 2-wire version) with display operates correctly only if both outputs are connected. Modular Humidity and Temperature Sensor | 9...
Fix the EE212 on the mounting plate with four M4x10 screws (included in the scope of supply). Insert the EE212 in the radiation shield and fix the mounting plate with two M4x10 screws on the radiation shield (included in the scope of supply).
5 Setup and Configuration The EE212 is ready to use and does not require any configuration by the user. The factory setup of EE212 corresponds to the type number ordered. Please refer to the datasheet at www.epluse.com/ee212. If needed, the user can change the factory setup with the help of the free EE-PCS Product Configuration Software and the USB configuration adapter (HA011066).
Analogue output configuration options: ▪ Measurand assignment ▪ Current and/or voltage scaling ▪ Unit selection ▪ Output value fixing (measured value or current/voltage value) Fig. 5 EE212 with current or voltage output (selectable) Modular Humidity and Temperature Sensor | 12...
Page 13
User Manual EE212 Fig. 6 EE212 with current output Digital output configuration options: ▪ Address ▪ Baud rate ▪ Parity ▪ Stop bits Fig. 7 EE212 digital interface communication settings Modular Humidity and Temperature Sensor | 13...
User Manual EE212 Display configuration options: ▪ Display layout ▪ Measurand selection ▪ Display appearance (backlight, brightness, contrast) Fig. 8 EE212 display configuration RS485 Digital Interface 5.2.1. Hardware Bus Termination For bus termination EE212 features an internal 120 Ω resistor which can be activated using the jumper on the electronics board (see also chapter 3.3 Electrical Connection for details).
Tab. 2 BACnet protocol settings PLEASE NOTE The recommended settings for multiple devices in a BACnet MS/TP network are 38 400, 8, none, 1. The EE212 PICS (Product Implementation Conformance Statement) is available on the E+E website at www.epluse.com/ee212. BACnet address and baud rate can be set via: ▪ EE-PCS Product Configuration Software and the USB configuration adapter HA011066.
Page 16
0x09 1) Register number starts from 1. 2) Protocol address starts from 0. 3) For Modbus protocol settings see Application Note Modbus AN0103 (available at www.epluse.com/ee212). Tab. 4 EE212 registers for device setup Modular Humidity and Temperature Sensor | 16...
°C, °F 0x134 1) The choice of measurement units (metric or non-metric) must be done according to the ordering guide, see EE212 datasheet. Switching from metric to non-metric or vice versa by using the EE-PCS is not possible. 2) Register number starts from 1 3) Register address starts from 0 4) Examples: For scale 100, the reading of 2550 means a value of 25.5.
The sensing module with rapidX technology allows a fast and tool-free exchange even during operation. There is no need to switch off the EE212 for this purpose (hot-swap). PLEASE NOTE Extreme working conditions such as highly polluted, aggressive and chemically contaminated environment might require periodical maintenance of the device.
The blue cable stiffener needs to be on the left side as shown in Fig. 11. NOTICE Disconnecting or connecting the display while the EE212 is powered may cause damages to EE212 and to the display. Always switch off the EE212 before exchanging the display Repairs Repairs may be carried out by the manufacturer only.
Sensing Module for EE212 Single packed EE212M-PK4 Multipackage / Tray EE212M-PK6 1) For replacement only, not for retrofitting an EE212 without display 2) Minimum order quantity: 10 pcs For accessories please refer to the EE212 datasheet at www.eplusee.com/ee212 and to the Accessories datasheet.
Page 22
Protection rating IP65 / NEMA 4X Electromagnetic compatibility EN 61326-1:2013 EN 61326-2-3:2013 Industrial environment FCC Part15 Class A ICES-003 (A) Conformity 1) For display operation with EE212-xA6 (4 - 20 mA, 2-wire) both outputs must be connected. Modular Humidity and Temperature Sensor | 22...
User Manual EE212 8 Conformity Declarations of Conformity E+E Elektronik Ges.m.b.H. hereby declares that the product complies with the respective regulations listed below: European directives and standards. UK statutory instruments and designated standards. Please refer to the product page at www.epluse.com/ee212...