R E V O L U T I O N W I R E L E S S ™
7 . t r O u B l e S H O O t i N G
Pr ob lem
Display does not come on
Weak stimulation with charged
batteries
Stimulation stops with charged
batteries
Stimulation weakens within
minutes of starting treatment
with charged batteries
Stimulation is uncomfortable
Stimulation is ineffective
Stimulation only felt on one
electrode
Po ss ible c ause
Low batteries
Electrodes dried out, lost their
adhesive power and have no
adequate connection to th skin.
Electrode placement
Poor electrode contact
Damaged or worn electrodes or
module leadwires
This is a normal body adaptive
process
Amplitude (intensity) is too high
Electrodes are too close together
Damaged or worn electrodes or
leadwires of modules
Ensure proper program is being
used
Improper electrode placement
Unknown
Improper electrode placement
S olution
Charge batteries
Replace electrode
Make the electrodes at least
2 inches apart
Reapply electrodes, secure firmly.
Electrodes must be a minimum
of 2 inches apart.
Replace
Increase the amplitude
(intensity) if required.
Decrease amplitude (intensity).
Reposition the electrodes.
Electrodes must be a minimum
of 2 inches apart.
Replace
A. Refer to section 6.3.1.2
Programs and Parameters for
a description of the Programs
B. Contact clinician if discomfort
persists.
Reposition electrodes. Electrodes
must be a minimum of 2 inches
apart.
Contact clinician.
A. Reposition electrodes.
Electrodes must be a
minimum of 2 inches apart.
B. Replace electrodes.
E N