Specifications
Performance Conditions
AWG7000 Series Arbitrary Waveform Generators Technical Reference
This section contains the specifications for the AWG7101, AWG7102,
AWG7051, and AWG7052 Arbitrary Waveform Generators.
All specifications are guaranteed unless noted as "typical". Typical specifications
are provided for your convenience but are not guaranteed. Specifications that are
marked with the n symbol are checked in the Performance Verification section
of this manual.
To meet specifications, following conditions must be met:
H The instrument must have been calibrated/adjusted at an ambient tempera-
ture between +20 _C and +30 _C.
H The instrument must have been operating continuously for 20 minutes within
the operating temperature range specified.
H The instrument must be in an environment where the temperature, altitude,
humidity, and vibration conditions are within the operating limits described
in these specifications.
1-1