Performance Tests
2-36
16. Press the Ch 1 On button on the AWG7000 to enable the channel 1 output.
17. Press the Run button on the AWG7000 to output the waveform.
18. Make the AWG7000 Low Level setting shown in the first row of Table 2-8
on page 2-35.
19. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_2.
20. Use the following formula to compensate the voltage for the 50 Ω BNC
terminator:
Marker_Low = (Term_R + 50) / (2 × Term_R) × Measured_voltage_2
21. Verify that the marker Low level falls within the limits given in Table 2-8 on
page 2-35.
22. Repeat steps 18 through 21 for each row in Table 2-8 on page 2-35.
23. Press the Ch 1 On button to disable the channel 1 output.
24. Move the SMA-BNC adapter from the Channel 1 Mkr 1 connector to the
Channel 1 Mkr 1 connector and move the 50 Ω SMA terminator from the
Channel 1 Mkr 1 connector to the Channel 1 Mkr 1 connector.
NOTE. For the Mkr 1 output, read marker_hi and marker_low as marker_low
and marker_hi respectively.
25. Repeat steps 8 through 22.
26. Repeat steps 8 through 25 for Channel 1 Mkr 2 and Mkr 2.
27. For the AWG7102 or AWG7052: Repeat the test for the Channel 2 marker
outputs.
28. Press the All Outputs On/Off button on the AWG7000 to turn off all the
outputs.
29. Disconnect the test setup.