8. Set the oscilloscope settings as indicated below:
Vertical scale
. . . . . . . .
Horizontal scale
Input coupling
. . . . . . .
Input impedance
CH 1 position
. . . . . . .
CH 2 position
. . . . . . .
CH 3 position
. . . . . . .
Trigger source
. . . . . . .
Trigger level
. . . . . . . .
Trigger slope
. . . . . . . .
Trigger mode
. . . . . . . .
9. Press the Factory Default button on the AWG7000.
10. Press the Ch1 Select button on the AWG7000.
11. On the AWG7000, load the sine_mk1_mk2 waveform as an output
waveform.
a. Press the File Open button or select File > Open File to open the Open
dialog box.
b. In the dialog box, navigate to the C:\Program Files\Tektronix\AWG\
System\PV directory, and then select the pv_awg7000.awg file. The
Waveform List window appears.
NOTE. If your instrument is the AWG7051 or AWG7052, a warning message is
displayed when you open the pv_awg7000.awg file. Press the OK button.
c. In the window, select (drag and drop) the sine_mk1_mk2 waveform on
the User Defined tab.
12. Press the Ch 1 On button on the AWG7000 to enable the cannel 1 output.
13. Press the Run button on the AWG7000 to output the waveform.
1 V/div (CH1, CH2, and CH3)
. . . . .
20 ns/div (for the AWG710x)
40 ns/div (for the AWG705x)
DC
. . . . .
50 Ω
+1 div (if necessary)
- -1 div (if necessary)
- -3 div (if necessary)