Tektronix AWG7000 Series Technical Reference page 73

Arbitrary waveform generators
Hide thumbs Also See for AWG7000 Series:
Table of Contents

Advertisement

AWG7000 Series Arbitrary Waveform Generators Technical Reference
11. On the oscilloscope, store the channel 1 waveform to Ref 1 as a reference
waveform.
12. On the AWG7000, set the Marker 1 delay value to 150 ps.
13. Use the oscilloscope to measure the delay time between the Ref 1 waveform
and channel 1 waveform at the 50% level.
14. Verify that the delay time is within the range of 92.5 ps to 207.5 ps.
15. Press the Ch 1 On button on the AWG7000 to disable the channel 1 output.
16. Move the SMA cable from the Channel 1 Mkr 1 connector to the Channel 1
Mkr 2 connector.
17. Move the SMA terminator from the Channel 1 Mkr 1 connector to the
Channel 1 Mkr 2 connector.
18. Press the Ch 1 On button on the AWG7000 to enable the channel 1 output.
19. On the oscilloscope, store the channel 1 waveform to Ref 1 as a reference
waveform.
20. On the AWG7000, set the Marker 2 delay value to 150 ps.
21. Repeat steps 13 and 14.
22. For the AWG7102 and AWG7052: Repeat the test for the Marker 1 and
Marker 2 of the Channel 2.
23. Press the All Outputs On/Off button on the AWG7000 to turn off all the
outputs.
24. Disconnect the test setup.