Tektronix AWG7000 Series Technical Reference page 35

Arbitrary waveform generators
Hide thumbs Also See for AWG7000 Series:
Table of Contents

Advertisement

Performance Verification Procedures
AWG7000 Series Arbitrary Waveform Generators Technical Reference
Two types of performance verification procedures can be performed on the
instrument: Brief Procedures and Performance Tests. You may not need to
perform all of these procedures, depending on what you want to accomplish.
H To rapidly confirm that the instrument functions and was adjusted properly,
perform Diagnostics and Calibration beginning on page 2-3.
Advantages: These procedures are quick to do and require no external
equipment or signal sources. These procedures perform extensive functional
and accuracy testing to provide high confidence that the instrument will
perform properly.
H To further check functionality, first perform Diagnostics and Calibration, and
then perform Functional Test beginning on page 2-5.
Advantages: The procedure requires minimal additional time to perform,
and requires minimal equipment. The procedure can be used when the
instrument is first received.
H If more extensive confirmation of performance is desired, complete the self
tests and functional test, and then do the Performance Tests beginning on
page 2-9.
Advantages: These procedures add direct checking of warranted specifica-
tions. These procedures require suitable test equipment and more time to
execute (refer to Equipment Required on page 2-9).
If you are not familiar with operating this instrument, refer to the online help or
the user information supplied with the instrument.
2-1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Awg7101Awg7102Awg7051Awg7052