NOTE: 1. If the system is unlikely to be used for a long period after the power is
turned OFF, turn OFF the main power switch on the power panel and then
unplug the power cable.
2. If the power cannot be turned OFF by following the normal procedures, 2D
mode images may not be displayed when the system power is turned ON
again. In this case, perform the following procedures:
 Shut down the system and turn OFF the main power switch on the
power panel on the rear of the system.
 Wait for at least 15 seconds, turn ON the main power switch on the
power panel, and then turn ON the power again.
3. After the main power has been turned OFF on the power panel, wait for at
least 15 seconds before turning the power ON again.
If the main power switch is turned ON immediately after the power has
been turned OFF, the system may malfunction.
4. The following message may be displayed during operation.
This message indicates that a problem has occurred with the patient data
handling functions (image storage, playback, or deletion) and that the
system has entered restricted operation mode to protect the data.
In restricted operation mode, do not use the Store, Exam Review, or
Patient Browser functions. If these functions are used, the patient data
handling functions cannot be performed. Reboot the system as soon as
the current examination is completed.
Note that the message can be cleared by clicking [Close] and scanning
can be performed normally.
No. 2B771-004EN*M