Turn the ring in the direction shown to tighten it.
(g) After the trackball is assembled, confirm the following:
 The ball and dial operate smoothly, there is no abnormal sound, and the
ball is not loose.
 The trackball works normally after the power is turned ON.
Cleaning the air filters
Check and clean the air filters to prevent overheating of the system or a reduction
in system performance/reliability due to clogging of the filters.
CAUTION:
Do not boot up the system while the air filters are removed.
The system may malfunction.
(a) Remove the air filters (3 locations).
(b) Remove all dust from the filters using a vacuum cleaner at a location away
from the system.
(c) Remount the air filters.
No. 2B771-004EN*M
20-7