Toshiba Aplio 500 TUS-A500 Operation Manual page 117

Diagnostic ultrasound system
Table of Contents

Advertisement

STC setting and automatic update function setting
(1) The STC slider positions when
for the adjustment ranges. The movement range over which adjustment is
possible depends on the position when
figures below.
Adjustment range when the
sliders are at the center
(2) The changed offset can be saved or reset using the following switch in the
[OTHER] menu on the touch panel:
:
:
(3) Quick scan is automatically updated when any changes are made to the STC
settings. Note that the automatic update function can be disabled from the
[Preset] menu (refer to the operation manual Applications volume,
subsection 22.2.1, for details).
If automatic updating is disabled, the changed STC settings are valid until
is pressed.
is pressed are set as the center values
Example of adjustment range
when the sliders are not at the
center
Saves the offset.
Returns the saved offset to the original setting.
No. 2B771-004EN*M
14-11
is pressed as illustrated in the