Programming The Equipment Configuration; Temperature Sensors; Safety Switch-Off („Timeout") - BARTEC BENKE HYGROPHIL H 4230-12 Ex Zone II Operating Instructions Manual

Table of Contents

Advertisement

5.2.4
EQUIP.CONF. ?
1/0
5.2.4.1
EQUIPM.-CONFIG
T1 extern ? 1/0
EQUIPM.-CONFIG
T internal ? 1/0
5.2.4.2
EQUIPM.-CONFIG
TIMEOUT ? 1/0
EQUIPM.-CONFIG
TIMEOUT [s] = 1200
Programming the equipment configura-
tion
0 or ENTER
Don't make any changes, go to the next menu (limit
monitoring)
Change the programming of the equipment configura-
1
tion

Temperature sensors

HYGROPHIL H 4230 contains two temperature measuring circuits for the
following temperature ranges:
The temperature measuring circuits are designed for the connection of
PT 100 temperature sensors and can be used as you wish.
T1
is an external temperature sensor. In order to measure temperature-
external
related humidity variables (RH, h) you must install the T1
sensor in the place where the humidity variable is to be measured.
Don't make any changes
ENTER
Switch on temperature sensor T1
1
Switch off temperature sensor T1
0
Temperature sensor T
internal
used to compensate the humid temperature (HT).
Don't make any changes
ENTER
Switch on temperature sensor T
1
Switch off temperature sensor T
0
The internal temperature sensor should always be switched on in
standard use of the device!
Safety switch-off („TIMEOUT")
You can program HYGROPHIL 4230 in such a way that it switches off fol-
lowing a defined time after a fault occurs (TIMEOUT).
You can restart the equipment following a „TIMEOUT" by pressing the
RESET key or by disconnecting the equipment from the mains and then
connecting it again.
If the equipment switches off automatically for safety reasons, you should
immediately attempt to locate the cause. If it is not possible to rectify the
fault straightaway, switch off the equipment and the compressed air supply.
0 or ENTER
Don't make any changes, go to the next menu (water
detector)
Change TIMEOUT programming
1
Don't make any changes
ENTER
Change the time for automatic switch-off: enter a new value in seconds us-
ing the keyboard, ENTER
(600...3599)
Operating Instructions HYGROPHIL® H 4230-12 BA 060207
T1
:
0...200 °C
external
T
:
0...100 °C.
internal
is installed permanently in the enclosure. It is
Programming
55
temperature
external
external
external
internal
internal