BARTEC BENKE HYGROPHIL H 4230-12 Ex Zone II Operating Instructions Manual page 108

Table of Contents

Advertisement

PROFIBUS
102
7
6
Free
Free
7
6
Free
Free
0
0
Default
DaAdr $57: EquipmentControlStatus
Format: Char
5
4
ExternWater
Free
Error
ExternWater: 1 = External level sensor reports „no water"
InternWater Error:
1 = Internal level control reports „no water"
If the corresponding inputs in EquipmentChConfig (DaAdr $4C) have not
been released, the status bits are left at '0'.
DaAdr $58: EquipmentChConfig
Format: Char
Before you can write this variable you must first change to standby mode.
5
4
External
Free
water
de-
tector
0
0
Releases the corresponding monitoring of the water level.
DaAdr $53: Limit Task Config
Format: Char [2]
Char [1] : 0x00 = Limit monitoring not active
0x01 = Limit monitoring active
Char [2] : Value for the measurand to be taken over.
0
TT
1
HT
2
SP
3
DP
4
T1
extern
5
T
intern
6
DT
7
RH
Operating Instructions HYGROPHIL® H 4230-12 BA 060207
3
2
Free
Free
3
2
Free
Free
0
0
8
VP
9
MH
10 SH
11 H
12 DVP
13 RH
extern
14 Vol%
1
0
InternWater
Free
Error
1
0
Internal wa-
Free
ter detector
1
0