Clear; Auto; Run/Stop; Single - Rigol DS1104Z-S Plus Quick Manual

Digital oscilloscope
Hide thumbs Also See for DS1104Z-S Plus:
Table of Contents

Advertisement

RIGOL

CLEAR

AUTO

Note:
Waveform auto setting function requires that the frequency of sine is no lower
than 41 Hz; the duty cycle should be greater than 1% and the amplitude must be at
least 20 mVpp for square. Otherwise, the Waveform auto setting function may be
invalid and the quick parameter measurement function displayed in the menu will
also be unavailable.

RUN/STOP

SINGLE

14
Press this key to clear all the waveforms on the screen. If the
oscilloscope is in "RUN" state, new waveforms will still be
displayed.
Press this key to enable the waveform auto setting function.
The oscilloscope will automatically adjust the vertical scale,
horizontal time base and trigger mode according to the input
signal to realize optimum waveform display.
Press this key to "RUN" or "STOP" waveform sampling. In the
"RUN" state, the key is illuminated in yellow. In the "STOP"
state, the key is illuminated in red.
Press this key to set the trigger mode to "Single". In single
trigger mode, press FORCE to generate a trigger signal
immediately.
DS1000Z Quick Guide