SDS-MW Analysis Kit

Sample Vial Setup

6
Place the vial in a room-temperature water bath to cool for 5 minutes before injection.
Sample Vial Setup
Before placing the 200 microliter sample vials (or micro vials) into the universal vials, ensure that
no bubbles are at the bottom of the micro vials. If bubbles exist, centrifuge the micro vials for 2
minutes at 1,000 g and repeat if necessary. Place a blue cap on the universal vial and ensure a good
seal. See
Figure
Place the universal vials into the 48-position inlet sample tray from positions A1 through C8.
Figure 1.1 Sample Vial Setup
1. Universal Cap
2. Micro Vial
1-8
1.1.
3. Universal Vial
4. Micro Vial inside Universal Vial
A51970AD