Write-In Of Continuous Word Data - Fuji Electric ZPA Instruction Manual

Infrared gas analyzer, communication functions modbus
Hide thumbs Also See for ZPA:
Table of Contents

Advertisement

6.4
Write-in of continuous word data [Function code:10
Function code
Max. word number write-in
in one message
10
H
(1)
Message composition
Command message composition (byte)
Station No.
Function code
Write-in start No.
(relative address)
Write-in word
number
Write-in byte number
First write-in word
data
Next write-in word
data
Last write-in word
data
CRC data
* Arrangement of write-in word data
MSB
Upper byte of contents of the first word data
Lower byte of contents of the first word data
Upper byte of contents of the next word data
Lower byte of contents of the next word data
Upper byte of contents of the last word data
Lower byte of contents of the last word data
(2)
Function explanation
Word data of continuous word number is written from write-in start address. Write-in word data are
transmitted from master station in the order of upper and lower bytes.
INZ-TN5A1190-E
Relative data address
64 words
Upper
Lower
Upper
1 to 64
Lower
Write-in word number×2
Upper
Lower
Upper
Lower
Upper
Lower
Lower
Upper
0000
-009D
40001-40158
H
H
Response message composition (byte)
LSB
]
H
Resister No.
Kind of data
User setting
Station No.
Function code
Write-in start No.
(relative address)
Write-in word
number
CRC data
Upper
Lower
Upper
Lower
Lower
Upper
21