Supplement To Address Map - Fuji Electric ZPA Instruction Manual

Infrared gas analyzer, communication functions modbus
Hide thumbs Also See for ZPA:
Table of Contents

Advertisement

7.3

Supplement to address map

* (a) Register No. 30062 to 30131 (error log)
Up to 14 errors logged can be read in the order from the latest to older ones.
The contents are as follows.
Error No.:
Error DAY:
Error HOUR:
Error MIN:
Error TARGET: No. of Ch where error occurred.
* (b) Register numbers 30181, 30182, 30183 (display information (1), (2), (3))
The display information is values for knowing the current display status of the instrument.
• Contents of values of display information (1) (status of each setting panel)
0: Measurement mode display (manual calibration display included)
1: Menu mode display
2: Changeover of range display
3: Calibration setting display
4: Alarm setting display
5: Automatic calibration setting display
6: Peak alarm setting display
7: Parameter setting display
8: Maintenance mode display
9: Factory mode display
10: Auto zero calibration setting display
• Contents of values of display information (2) (status at manual calibration)
0: Measurement mode display
4: Channel selection display at manual zero calibration
5: Zero calibration wait display at manual zero calibration
6: Zero calibration in progress display at manual zero calibration
7: Channel selection display at manual span calibration
8: Span calibration wait display at manual span calibration
9: Span calibration in progress display at manual span calibration
10: Error contents indication display
• Contents of values of display information (3)
Correspond to value of Ch (channel) from which the measurement mode is displayed.
Top channel number minus 1 is stored.
* (c) Register number 30189 (manual calibration channel)
Data for knowing a channel (component) which the cursor is positioned at to carry out a calibration at
when a manual calibration is under way.
Channel number to carry out a calibration at minus 1 is stored.
38
No. of produced error. Stored value is error number minus 1.
Date
Indicates at what o'clock error occurred.
Indicates at what minutes error occurred.
Ch No. minus 1 is stored.
0 at error No. 3 or 10.
INZ-TN5A1190-E