TROUBLESHOOTING

If the communication is unavailable, check the following items.
Whether all devices related to communication are turned on.
Whether connections are correct.
Whether the number of connected instruments and connection distance are as specified
Whether communication conditions coincide between the master station (host computer) and slave stations
(instrument)
Transmission speed
Data length
Stop bit
Parity
Whether send/receive signal timing conforms to Section 5.6 in this manual.
Whether the station No. designated as send destination by the master station coincides with the station No. of
the connected instrument.
Whether more than one instrument connected on the same transmission line shares the same station No.
Whether the station No. of instruments is set at other than 0.
If it's 0, the communication function does not work.
INZ-TN5A1190-E
:
38400bps
:
8 bits
:
1 bit
:
None