mA Read Power Current Measurement
In this mode, the PCE-TTC 30 works as Loop Power Supply
the
current.
Voltage Measurement
The PCE-TTC 30
is capable of voltage Measurement with two voltage measurement ranges.
The following picture displays the connection for Voltage Measurement for different
mode.
TC Terminals
Terminal for measuring and simulating thermocouples and mV. This terminal accepts a
miniature polarized thermocouple plug with flat, inline blades spaced 7.9 mm (0.312 in)
center to center.
TC Type
-200.0 ... 1000.0 °C
E
-200.0 ... 1200.0 °C
J
-200.0 ... 1372.0 °C
K
-200.0 ... 400.0 °C
T
450.0 ... 1800.0 °C
B
0.0 ... 1750.0 °C
R
0 ... 1750.0 °C
S
-200.0 ... 1300.0 °C
N
-10 ... 80 mV
mV
-10 ... 250 mV
The PCE-TTC 30 supports measurement and simulation of Thermocouple and mV.
TC Terminal (Measure and Source)
Range
Resolution
© PCE Instruments
while at the same time measuring
Display
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.001 mV
+0.02 % of reading + 2µV
0.01mV
+0.02 % of reading + 0.02mV
5
Accuracy
± 0.3 °C
± 0.3 °C
± 0.3 °C
± 0.3 °C
± 0.5 °C
± 0.5 °C
± 0.5 °C
± 0.3 °C