Installation; Location; Wiring - Thermo Scientific TSX Series Installation And Operation Manual

Ultra low temperature freezers
Table of Contents

Advertisement

Installation

7 Installation

7.1 Location

7.2 Wiring

8
Installation and Operation Manual
WARNING: Do not exceed the electrical rating printed on the data plate
located on the lower left side of the unit.
Install the unit in a level area free from vibration with a minimum of eight inches
(20 cm) of space on the top and sides, six inches (15 cm) in back. (Refer to
Section 7.3 for further instructions on leveling cabinets.) Allow enough clearance
so that door can swing open at least 85°.
The rear spacing posts provided with the freezer can be used to ensure proper
clearance. To install the spacing posts, screw them into the back in the rear deck area.
Do not position the equipment in direct sunlight or near heating diffusers,
radiators, or other sources of heat. The ambient temperature range at the location
must be 59 to 90°F (15 to 32°C).
CAUTION: Connect the equipment to the correct power source. Incorrect
voltage can result in severe damage to the equipment.
CAUTION: For personal safety and trouble-free operation, this unit must be
properly grounded before it is used. Failure to ground the equipment may cause
personal injury or damage to the equipment. Always conform to the National
Electrical Code and local codes. Do not connect the unit to overloaded power
lines.
CAUTION: Do not position the unit in a way that impedes access to the
disconnecting device or circuit breaker in the back of the unit.
CAUTION: Always connect the freezer to a dedicated (separate) circuit. Each
freezer is equipped with a service cord and plug designed to connect it to a
power outlet which delivers the correct voltage. Supply voltage must be within
+10% or -10% of the freezer rated voltage. If cord becomes damaged , replace
with a properly rated power supply cord 3-G 1.5mm², CEE 7/7, 16A / 250 V .
CAUTION: Never cut the grounding prong from the service cord plug. If the
prong is removed, the warranty is invalidated.