Hartmann Veroval Instructions For Use Manual page 98

Upper arm blood pressure monitor
Hide thumbs Also See for Veroval:
Table of Contents

Advertisement

Available languages

Available languages

English
GB
Table 4
For all NON-LIFE-SUPPORT medical-electrical (ME) DEVICES and ME SYSTEMS
Recommended separation distances between portable and mobile RF communica-
tions equipment and the Veroval
The Veroval
upper arm blood pressure monitor is intended for use in an electromagnetic
®
environment in which radiated RF disturbances are controlled. The customer or user of the
Veroval
upper arm blood pressure monitor can help prevent electromagnetic interference
®
by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Veroval
mended below, according to the maximum output power of the communications equip-
ment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be determined using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
upper arm blood pressure monitor
®
upper arm blood pressure monitor, as recom-
®
Separation distance according to transmitter frequency (m)
80 MHz to 800 MHz
0.1167
0.3689
1.1667
3.6893
11.6667
98
800 MHz to 2.5 GHz
0.2334
0.7378
2.3334
7.3786
23.3334

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents