Never inflate the cuff when it is not properly applied to the upper arm.
Use the monitor only with the approved upper arm cuff. The monitor may oth-
erwise sustain external or internal damage.
The cuff tube may only be removed from the unit by pulling on the corre-
sponding connector plug. Never pull on the tube itself!
Do not expose the monitor to extreme temperatures, humidity, dust or direct
sunlight because this may cause it to malfunction.
Keep the packaging, batteries and monitor out of the reach of children.
Please comply with the storage and operating conditions defined in Chapter
13 Technical data. Storage or use outside the specified temperature and
humidity range can affect the accuracy of the measurement or the function of
the monitor.
Do not use the monitor near strong electromagnetic fields and keep away from
radio equipment or mobile phones. Portable and mobile high-frequency and
communication devices, such as telephones and mobile phones, can impair the
functionality of this electronic medical device.
Instructions for the metrological check
Every Veroval
monitor has been carefully tested by HARTMANN for measure-
®
ment accuracy and has been developed with a view to a long service life. We
recommend carrying out a metrological check every 2 years for devices in pro-
fessional use, for example, in pharmacies, medical practices or hospitals. Please
also observe the national regulations specified by the legislator. Metrological
checks may only be carried out by the competent authorities or authorised main-
tenance providers against cost reimbursement.
Instructions for calibration mode
A functional check of the monitor can be carried out on people or using a suit-
able simulator. A metrological check involves testing for leaks in the pressure
system and possible deviations in the pressure reading. Remove at least one
battery to switch to calibration mode. Now hold the START/STOP button down
and re-insert the battery. Release the button after a few seconds and, after a few
moments, 'FA(' will appear on the display. 'PSr' and '01' will then appear.
Press the START/STOP button briefly. '000' and '0' will appear on the display.
Instructions on the metrological check will be supplied by HARTMANN upon
request to the competent authorities or authorised maintenance providers.