Creating a Data Set

To fully utilise the Alaris PK Plus Syringe Pump a Data Set will need to be developed, reviewed, approved, released, uploaded and
verified according to the following process. Refer to the Alaris PK Editor Software Directions for Use (1000CH00016) for further details
and operating precautions.
1. Create Master Lists (Using Alaris PK Editor Software)
• Master Drugs*
• Alaris PK Plus Syringe Library
2. Create Profile (Using Alaris PK Editor Software)
• Profile Drugs*
• Pump Configuration**
3. Review, Approve and Release (Using Alaris PK Editor Software)
• Review and Approve
• Release
4. Upload Data Set to Alaris PK Plus Syringe Pump (Using Alaris PK Editor Transfer Tool)
5. Verify Data Set Upload
• First or Individual Pump Verification
• Subsequent Pump Verification
6. Switch the Pump on and verify that the start-up splash screen displays the correct data set name and version. The Pump is now
ready to use.
w
*Drug parameters have to be in accordance to local protocols and prescribed information.
Data set transfers should only be performed by qualified technical personnel.
** See important note in Configured Options section.
A list of drug names and standard concentrations. These may be for TIVA use or
may have an associated PK/PD model for TCI use.
Configure syringes enabled for use
Drugs and concentrations for this profile with defaults, minimum and
maximum limits and targets and occlusion level.
Pump configuration settings and general options.
Entire Data Set Report to be printed, reviewed and signed as proof of approval
by an authorised person according to Hospital protocol. Signed printout to be
kept safe for use during verification procedure.
Data Set status to be promoted to Released (password is required).
On completion of upload record CRC (Cyclic Redundancy Check) number
shown on the Alaris PK Plus Syringe Pump.
Download the Data Set from the Pump using the Alaris PK Verification Tool.
Compare Data Set downloaded with the approved signed Data Set printout.
Reviewer should sign the printout and also record the CRC number on the
printout as record.
On subsequent uploads of the Data Set compare CRC number on the Pump
with CRC number recorded on First Pump Verification.
1000DF00741 Issue 5
Alaris™ PK Plus Syringe Pump
7/52
Creating a Data Set