Cleaning and Storage
Before the transfer of the Pump to a new patient and periodically during the use, clean the Pump by wiping over with a lint-free cloth
lightly dampened with warm water and a standard disinfectant / detergent solution.
Do not use the following disinfectant types:
• Disinfectants which are known to be corrosive to metals must not be used, these include:
• NaDcc (such as Presept),
• Hypochlorites (such as Chlorasol),
• Aldehydes (such as Cidex),
• Cationic Surfactants >1% (such as Benzalkonium Chloride).
• Mixture of Alcohol & Chemicals with Cationic surfactants >1% Chlorohydrocarbons (such as Amberclens)
• Use of Iodine (such as Betadine) will cause surface discoloration.
• Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Recommended cleaners are:
Brand
Concentration
Hibiscrub
20% (v/v)
Virkon
1% (w/v)
The following products were tested and are acceptable for use on the Pump if used in accordance with the specified manufacturer's
guidelines.
• Warm soapy water
• Mild detergent in water (e.g. Young's Hospec)
• 40% Isopropyl Alcohol in water
• Chlor-Clean
• Hibiscrub
• Tristel Fuse sachets
• Tristel Trio wipes system
• Tuffie 5 wipe
• Virkon Disinfectant
w
Before cleaning always switch off and disconnect from the AC Power Supply. Never allow liquid to enter the casing
and avoid excess fluid build up on the Pump. Do not use aggressive cleaning agents as these may damage the exterior
surface of the Pump. Do not steam autoclave, ethylene oxide sterilise or immerse this Pump in any fluid.
If the Pump has visible cracks or damage to the case do not clean and immediately take it out of service for
examination by Qualified Service Personnel.
The syringe and extension sets are disposable single use items and should be discarded after use according to their manufacturers'
instructions.
If the Pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged. Store in a clean, dry
atmosphere at room temperature and, if available, employ the original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in the Technical Service Manual and ensure that the internal
battery is fully charged.
Disposal
Information on Disposal for Users of Waste Electrical and Electronic Equipment
U
This
symbol on the product and/or accompanying documents means that used electrical and electronic products should not be
mixed with household waste.
If you wish to discard electrical and electronic equipment, please contact your CareFusion affiliate office or distributor for further
information.
Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and
the environment which could otherwise arise from inappropriate waste handling.
Information on Disposal in Countries outside the European Union
U
This
symbol is only valid in the European Union. The product should be disposed of taking environmental factors into consideration.
To ensure no risk or hazard, remove the internal rechargeable battery and the Nickel Metal Hydride battery from the control board and
dispose of as outlined by the local country regulations. All other components can be safely disposed of as per local regulations.
1000DF00741 Issue 5
Alaris™ PK Plus Syringe Pump
42/52
Maintenance