Alaris PK Editor Software - Pump Configuration
The following options are configurable via the Alaris PK Editor Software (PC based), see the Alaris PK Editor Software Directions for Use
(1000CH00016) for details on how to alter the profile configurations.
General Pump Configurations
AC Fail Warning
Audio Volume
Auto Night Mode
Battery Icon
Callback Time
Event Log
Drug Override Mode
Pressure Default
Pressure Display
Purge Rate
Purge Volume Max
Purge Syringe Prompt
1
Bolus
1
Bolus Rate Default
1
Bolus Volume Default
KVO
KVO Rate
Near End of Infusion Time
End of Infusion %
2
Weight Default
2
Weight Minimum
2
Weight Maximum
2
Age Default
2
Age Minimum
2
Age Maximum
w
The approved data set contains configurable option values per profile.
1
The bolus configurations are used only when the Alaris PK Plus Syringe Pump is being used in ml/h mode. If a drug is selected then the
drug's own configuration settings are used.
2
Although a default and Soft Limits can be set for age and weight, the actual selectable range may be limited by the drug and model
chosen.
The AC Power Failure Alarm can be set to sound or be silent if the AC power is disconnected.
The audio alarm volume of the Pump (high, medium or low).
Main Display (Backlight) dims between hours 21:00 and 06:00.
Indicator displaying the remaining estimated battery capacity.
Adjusts the length of time before the Pump sounds the Call Back alarm.
The Event Log can be set to be displayed on the main display. Events are still recorded in the Event
Log if disabled.
Always - Any changes made to the dose rate or target concentration that are outside the editor Soft
Alerts will require confirmation before starting infusion.
Smart - Confirmation of setting will be required on the first dose rate or target concentration set
outside the editor Soft Alerts. Any subsequent changes will not require confirmation until after the
dose rate or target concentration has been confirmed inside the editor Soft Alerts. Additionally, any
changes in dose rate or target concentration from above a Soft Alert Max to below a Soft Alert Min
or from below a Soft Alert Min to above a Soft Alert Max will also need to be confirmed.
The default occlusion pressure alarm level.
Sets whether the Pressure Information is available on the main display.
The rate used during purge operation.
The maximum permissible purge volume.
Feature which prompts the user to purge the extension set prior to the start of the infusion.
Bolus feature can be set to HANDS ON or HANDS FREE.
The default bolus rate.
The default bolus volume.
Allows the enabling or disabling of Keep Vein Open (KVO) at End of Infusion (EOI).
Sets the KVO rate at which the Pump will operate when EOI is reached.
Sets the Near End of Infusion warning time as time left to End of Infusion.
Sets the End of Infusion point as a percentage of syringe volume.
The patient default weight in kg.
The minimum patient weight in kg. This is a Soft Alert and can be overridden.
The maximum patient weight in kg. This is a Soft Alert and can be overridden.
The default patient age in years.
The minimum age in years. This is a Soft Alert and can be overridden.
The maximum age in years. This is a Soft Alert and can be overridden.