CareFusion Alaris PK Plus MK4 8005PK201 Directions For Use Manual page 35

Syringe pump
Table of Contents

Advertisement

End Of Infusion (EOI) Alarm
0.1% - 5% of syringe volume.
Maximum Pumping Pressure Limit
Highest alarm level 1000mmHg (nominal at L-10)
Occlusion Accuracy without pressure set (% of full scale)*
L-0
approx. 50mmHg
Temp. 23°C
±18%
* - Using most common 50ml syringes under normal conditions (95% confidence / 95% of pumps).
System Accuracy (continuous ml/h and TIVA)
Volumetric Mean +/- 2% (nominal).
• Derating - Temperature +/- 0.5% (5 - 40ºC), High Rates +/-2.0% (rates > syringe volume/h eg. >50ml/h in a 50ml syringe.)
w
System accuracy is +/-2% typical by volume as measured using the trumpet curve test method defined in
EN/IEC60601-2-24 at rates of 1.0ml/h (23ºC) and above when the Pump is used with the recommended syringes.
Caution: Infusion volume accuracy may be compromised at rates below 1.0ml/h. Differences in factors such as size
and plunger force in recognised syringes can cause variations in accuracy and trumpet curves. See also 'trumpet
curves' section in this manual.
Electrical Classification
Class I product. Continuous Mode Operation, Transportable.
Battery Specifications
Rechargeable sealed NiMH. Automatically charges when the Pump is connected to AC power.
Mean Time To Power Down from fully charged @ 5ml/h and 23°C ± 2°C under normal conditions is 6 hours*.
*95% lower confidence interval of 5 hours 50 minutes
Charging takes 2½ hours from discharge to 90% charge.
In TCI mode, a fully charged battery allows at least one full syringe to be infused.
Memory Retention
The electronic memory of the Pump will be retained for at least 6 months when not powered up.
Fuse Type
2 x T 1.25L250V
AC Power Supply
115 - 230VAC, 50 - 60Hz, 30VA (under maximum charging conditions) 10VA (minimum).
Dimensions
310 mm (w) x 121 mm (h) x 200 mm (d).
Weight
2.4 kg (excluding power cable).
Protection against fluid ingress
IP32 - Protected against direct sprays of water up to 15° from vertical and protected against solid objects greater than 2.5mm.
Note: IP33 applies if mains retainer kit, part number 1000SP01294, is fitted.
Pressure mmHg
L-3
approx. 300mmHg
approx. 500mmHg
±21%
1000DF00741 Issue 5
L-5
L-10
approx. 1000mmHg
±23%
±28%
34/52
Alaris™ PK Plus Syringe Pump
Specifications

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents