Hazards
An explosion hazard exists if the Pump is used in the presence of flammable anaesthetics. Exercise care to
locate the Pump away from any such hazardous sources.
A
Dangerous Voltage: An electrical shock hazard exists if the Pump's casing is opened or removed. Refer all
servicing to Qualified Service Personnel.
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If
the integrity of the external protective conductor in the installation or its arrangement is in doubt, the
Pump should be operated from the battery.
Do not open the RS232/Nurse Call protective covering when not in use. Electrostatic discharge (ESD)
V
precautions are required when connecting RS232/Nurse Call. Touching the pins of the connectors may
result in ESD protection failure. It is recommended that all actions must be taken by appropriately trained
personnel..
If the Pump is dropped, subjected to excessive moisture, fluid spillage, humidity or high temperature, or
otherwise suspected to have been damaged, remove it from service for inspection by Qualified Service
L
Personnel. When transporting or storing the Pump, use original packaging where possible, and adhere
to temperature, humidity and pressure ranges stated in the 'Specifications' section and on the outer
packaging.
The embedded Pump software incorporates limits and Pump configuration parameters. Qualified
personnel must ensure the appropriateness of the limits, the compatibility of the drugs, and the
performance of each Pump, as part of the overall infusion. Potential hazards include drug interactions,
and inappropriate delivery rates and pressure alarms.
Warning: Alaris Syringe Pumps should not be modified or altered in any way, except where explicitly
directed or authorised by CareFusion. Any use of Alaris Syringe Pumps which have been altered or
modified otherwise than in strict application of directions provided by CareFusion, is at your sole risk, and
CareFusion does not provide any warranty for or endorsement on any Alaris Syringe Pump that has been
so modified or altered. CareFusion product warranty shall not apply in the event the Alaris Syringe Pump
has suffered damage or premature wear, or malfunctions or otherwise operates incorrectly, as a result of
unauthorised modification or alteration of the Alaris Syringe Pump.
1000DF00741 Issue 5
Alaris™ PK Plus Syringe Pump