Specifications

Electrical Protection
Class I, Type CF (Defibrillation-proof )
Electrical/Mechanical Safety
Complies with IEC/EN60601-1 and IEC/EN60601-2-24
Potential Equalisation Conductor
The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Pump and the potential
equalisation busbar of the electrical installation . To use the Potential Equalisation Connector, connect the Potential Equalisation
Connector on the Pump to the potential equalisation busbar of the electrical installation .
Electro Magnetic Compatibility (EMC)
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24
Electrical Safety
Typical Earth Leakage Current (Normal Condition) = 78µA*
Typical Enclosure Leakage Current (Normal Condition) < 1µA*
Typical Protective Earth Resistance = 32mOhms (excluding power cable)
The above measurements are for guidance only, IEC/EN60601-1 limits are defined below:
Earth Leakage Current (Normal Condition) <= 500µA
Enclosure Leakage Current (Normal Condition) <= 100µA
Protective Earth Resistance <= 200mOhms (power cable)
* measured at 230 + 10% VAC, 60Hz
Classification
Continuous mode of operation, Portable Equipment
AC Power Supply
100 - 230VAC, 50 - 60Hz, 60VA (Maximum)
Dimensions
148mm (w) x 225mm (h) x 158mm (d) . Weight: approximately 2 .6kg (excluding power cable) .
Protection Against Fluid Ingress
IPX3 - Protected against direct sprays up to 60° from the vertical
Environmental Specifications
Condition
Temperature
Humidity
Atmospheric Pressure
*Non condensing
Battery Specifications
Rechargeable NiMH (Nickel Metal Hydride) . Automatically charges when the Pump is connected to AC power .
• Battery Run Time - For a 24 hour battery charge time, the Pump at 25ml/h will have a Mean Time To Battery Empty of 5 .5 hours .
• Battery Charging - 2 .5 hours to 95% .
Memory Retention
The electronic memory of the Pump will be retained for at least 2 years with normal use .
Operating
+5°C - +40°C
20% - 90%*
700hPa - 1060hPa
1000DF00660 Issue 9
Alaris™ VP Plus Guardrails™ Volumetric Pump
Transport and Storage
-20°C - +50°C
15% - 95%*
500hPa - 1060hPa