Priming The Infusion Set; Start Infusion - CareFusion Alaris VP Plus Guardrails Directions For Use Manual

Volumetric pump
Table of Contents

Advertisement

8 . Clear VOLUME infused , if required . This is recommended for a new patient or when a new infusion is set-up .
9 . Enter VTBI, if required, by selecting VTBI softkey on main display . Set VTBI by using the BAGS option and/or
press OK to confirm . Select the END RATE using the
softkey to confirm .
10 . Enter or adjust the RATE, if necessary, using the

Priming the Infusion Set

• Always prime the infusion set prior to connecting to the patient.
A
• The prime rate and prime volume max are configured in the data set via the Alaris Editor.
• The Pump will not prime if rate lock has been activated.
• During PRIME, the downstream occlusion alarm limit is temporarily increased to its maximum.
i
The
button allows the delivery of a limited volume of fluid in order to prime the infusion set prior to being connected to a patient .
i
11 . Press
button to display the PRIME screen .
12 . Press and hold the flashing PRIME softkey until the fluid flows and the priming of the infusion set is complete . The volume used
during priming will be displayed, but it is not added to the volume infused .
13 . When priming is complete release the PRIME softkey .

Start Infusion

14 . Connect the infusion set to the patient access device .
b
15 . Press
button to start the infusion . INFUSING will be displayed .
NOTE: The green run LED will flash to show that the Pump is infusing .
If the infusion requires to be stopped immediately, the following actions may be applied:
A
h
• press the
• close the roller clamp and then open the door.
f
f
button, recommended action.
1000DF00660 Issue 9
Alaris™ VP Plus Guardrails™ Volumetric Pump
keys to scroll through the on-screen choices and press the OK
keys .
17/48
Starting the Infusion