Infusion Specifications -
Maximum infusion rate can be set as part of the configuration.
0.1ml/h - 150ml/h
0.1ml/h - 300ml/h
0.1ml/h - 600ml/h
0.1ml/h - 900ml/h
0.1ml/h - 1200ml/h
The Volume Infused range is 0.0ml - 9990ml.
Bolus Specifications -
Maximum Bolus rates can be set as part of the configuration. Bolus
rates are user adjustable, in increments of 10ml/h.
10 ml/h - 150ml/h
10 ml/h - 300ml/h
10 ml/h - 600ml/h
10 ml/h - 900ml/h
10 ml/h - 1200ml/h
The bolus volume limit can be set as part of the configuration.
Minimum: 0.5ml; maximum 25.0ml
Increments of 0.1ml; default 5.0ml
During BOLUS the pressure limit alarms are temporarily increased to
their maximum level.
Critical Volume -
The bolus which can occur in the event of a single internal fault
condition with a 50 ml syringe is :
Maximum Overinfusion - 0.87ml
Purge Specifications -
The purge rate is limited to the maximum rate for the syringe and can
be set as part of the configuration.
100ml/h - 500ml/h.
The purge volume range is 0.5ml - 5ml.
During PURGE the pressure limit alarms are temporarily increased to
their maximum level.
Keep Vein Open (KVO) Rate -
0.1 ml/h - 2.5ml/h.
End Of Syringe Rate -
Stop, KVO (0.1ml/h to 2.5ml/h), or set rate if lower than KVO.
Volume To Be Infused (VTBI) -
0.1ml - 100ml, 1min - 24h
VTBI Complete Rate -
Stop, KVO (0.1ml/h to 2.5ml/h), set rate if lower than KVO or continue
at set rate.
Near End Of Infusion Alarm -
1min - 15min to end of infusion, or 10% of syringe volume, whichever
is smaller.
End Of Infusion (EOI) Alarm -
0.1% - 5% of syringe volume
Electrical Classification -
Class I product. Continuous Mode Operation, Transportable
Maximum Pumping Pressure Limit -
Highest alarm level 1000mmHg (nominal at L-10)
Occlusion Accuracy without pressure set (% of full scale)* -
L-0
approx.
50 mmHg
±18%
Temp. 23°C
Occlusion Accuracy with pressure set (% of full scale)* -
0
±2%
Temp. 23°C
±4%
Temp. 5°C-40°C
* - Using most common 50ml syringes under normal conditions
(95% confidence / 95% of pumps).
5ml syringes
10ml syringes
20ml syringes
30ml syringes
50ml syringes
5ml syringes
10ml syringes
20ml syringes
30ml syringes
50ml syringes
Pressure mmHg
L-3
L-5
L-10
approx.
approx.
approx.
300 mmHg 500 mmHg
1000 mmHg
±21%
±23%
±28%
Pressure mmHg
25
500
±4%
±5%
±6%
±7%
±7%
±10%
1000DF00333 Issue 6

Specifications

System Accuracy -
Volumetric Mean +/- 2% (nominal).
Derating -
Important: System accuracy is +/-2% typical by volume as measured
using the trumpet curve test method defined in EN/IEC60601-2-24 at
rates of 1.0ml/h (23ºC) and above when the pump is used with the
recommended syringes. Caution: Infusion volume accuracy may be
compromised at rates below 1.0ml/h. Differences in factors such as
size and plunger force in compatible syringes can cause variations
in accuracy and trumpet curves. See also trumpet curves section in
this manual.
®
Guardrails
Data Set Specification -
A maximum of 10 profiles can be set with a maximum of 100 drugs per
profile. See Guardrails
Battery Specifications -
Rechargeable sealed NiMH. Automatically charges when the pump is
connected to AC power.
Mean Time To Battery Empty from fully charged @ 5ml/h & 20°C under
normal conditions is 6 hours*
*95% lower confidence interval of 5 hours 50 minutes
Charging takes 2½ hours from discharge to 90% charge.
Memory Retention -
The electronic memory of the pump will be retained for more than 6
months when not powered up.
Fuse Type -
2 x T 1.25A, slow blowing.
AC Power Supply -
115 - 230VAC, 50 - 60Hz, 20VA (nominal).
Dimensions -
335 mm (w) x 121 mm (h) x 200 mm (d). Weight: 2.7 kg (excluding
power cable).
Protection against fluid ingress -
IPX1 - Protected against vertically falling drops of water.
Alarm Conditions -
Drive Disengaged
Check Syringe
Near End Of Infusion
VTBI Done
Internal Malfunction
Pressure Disc Out
Dose Would Exceed
Dose Under
Bolus Dose Under
Concentration not Permitted
Rate not Permitted
Environmental Specifications -
Operating Temperature
Operating Relative Humidity
Operating Atmospheric Pressure
Transport & Storage Temperature
Transport & Storage Relative Humidity
Transport & Storage Atmospheric Pressure
Electrical/Mechanical Safety -
Complies with EN/IEC60601-1 and EN/IEC60601-2-24.
1000
EMC -
Complies with EN/IEC60601-1-2 and EN/IEC60601-2-24.
Temperature +/- 0.5% (5 - 40ºC)
High Rates +/-2.0% (rates > syringe volume/h eg.
>50ml/h in a 50ml syringe.)
®
Editor Software DFU for more details.
Occlusion / Line Occlusion
Battery Low / Battery Empty
End of Infusion
AC Power Fail
Attention (Nurse Callback)
Titration not confirmed
Dose not Permitted
Bolus Dose Over
Bolus Dose not Permitted
Weight Outside Limit
26/38
+5°C - +40°C
20% - 90%
700hPa - 1060hPa
-30°C - +50°C
10% - 95%
500hPa - 1060hPa