System Accuracy
Rate Accuracy is ±5%, achieved under nominal conditions
Rate Accuracy is ±10%, achieved under low flow conditions
Variations of container head height, outlet back pressure or any combination of these can affect mean rate accuracy.
A
Factors that can influence container head height and outlet back pressure are: Flow rate, catheter length and inner
diameter, administration set length and inner diameter, container fill level and pump level, IV solution viscosity, and
IV solution temperature. The Notes under this section of the Directions For Use provide quantitative data on the
influences of intake and outlet pressures on mean flow rate.
Downstream Occlusion Alarm Pressure
Occlusion pressure is adjustable from 1mmHg to 800mmHg in 1mmHg increments .
Occlusion Accuracy achieved under nominal conditions
The downstream occlusion pressure accuracy can vary dependent upon temperature, with the table below showing the accuracy as %
Full Scale Deflection .
1
23°C
±2%
5°C
±4%
40°C
±4%
Note: The Pump does not display negative pressure values .
Maximum Pumping Pressure
880mmHg
Post Downstream Occlusion Bolus
Bolus volume generated at 25ml/h when the minimum downstream occlusion alarm limit is reached <0 .06ml
Bolus volume generated at 25ml/h when the maximum downstream occlusion alarm limit is reached <0 .70ml
Bolus Volume Accuracy
Typical: -1 .9%, Max: 4 .1%, Min: -10 .1% 1ml at 10ml/h
Typical: 0 .4%, Max: 1 .8%, Min: -1 .4% 100ml at 1200ml/h
Typical Time for Activation of the Downstream Occlusion Alarm
Downstream Occlusion Alarm Pressure
1mmHg
1ml/h
100 sec
25ml/h
11 sec
100ml/h
2 sec
Administering a Bolus
Parameter
Bolus Rate
Bolus Volume Displayed
Starting the Infusion/Set up
Infusion Parameter
Infusion Rate
VTBI Primary
VI (Total)
Air Sensor
Integral Ultrasonic Sensor .
Critical Volume
The maximum volume infused following a single fault condition is for rates ≤10ml/h: 0 .25ml, rates ≤100ml/h: 0 .5ml, rates >100ml/h: 2ml
The Alaris Safety Clamp
Set based, Pump activated Safety Clamp Device to prevent free flow
1a,
Pressure (mmHg)
25
500
±4%
±5%
±7%
±7%
±7%
±7%
30mmHg
165 sec
13 sec
4 sec
Range
10 - 1200ml/h in steps of 10ml/h
0 .0ml - 100 .0ml in steps of 0 .1ml
Range
0 .1 - 99 .9ml/h in steps of 0 .1ml/h
100 - 999ml/h in steps of 1ml/h
1000 - 1200ml/h in steps of 10ml/h
(0 - OFF), 1 - 9999ml
0 .1 - 9999ml
1000DF00660 Issue 9
Alaris™ VP Plus Guardrails™ Volumetric Pump
1a,2
1b,2
800
±6%
±10%
±10%
800mmHg
55 min
100 sec
20 sec
36/48