Trumpet and Flow Rate Curves
In this Pump, as with all infusion systems, the action of the pumping mechanism and variations cause short-term fluctuations in rate
accuracy .
The following curves show typical performance of the system in two ways: 1) the accuracy of fluid delivery over various time periods is
measured (trumpet curves), and 2) the delay in onset of fluid flow when infusion commences (start-up curves) .
Trumpet curves are named for their characteristic shape . They display discrete data averaged over particular time periods or 'observation
windows' , not continuous data versus operating time . Over long observation windows, short term fluctuations have little effect on
accuracy as represented by the flat part of the curve . As the observation window is reduced, short term fluctuations have greater effects
as represented by the mouth of the trumpet .
Knowledge of system accuracy over various observation windows may be of interest when certain drugs are being administered . Short
term fluctuations in rate accuracy may have clinical impact depending on the half-life of the drug being infused and the degree of inter
vascular integration, the therapeutic effect cannot be determined from the trumpet curves alone .
The start-up curves represent continuous flow versus operating time for two hours from the start of the infusion . They exhibit the delay
in onset of delivery due to mechanical compliance and provide a visual representation of uniformity . Trumpet curves are derived from
the second hour of this data . Tests performed per IEC/EN60601-2-24 standard .
Start-up Graph at 1 .0ml/h (Initial Period)
2.00
1.80
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
0
20
40
Time (minutes)
Trumpet Graph at 1 .0ml/h (Initial Period)
30
25
20
15
10
5
0
-5
-10
-15
-20
-25
-30
0
5
10
Observation Window (minutes)
Maximum Error
Minimum Error
60
80
100
120
15
20
25
30
Overall Mean = -0 .8%
1000DF00660 Issue 9
Alaris™ VP Plus Guardrails™ Volumetric Pump
Trumpet Graph at 1 .0ml/h (After 24 Hours)
30
25
20
15
10
5
0
-5
-10
-15
-20
-25
-30
0
5
10
Observation Window (minutes)
Maximum Error
Trumpet Graph at 1 .0ml/h (After 72 Hours)
30
25
20
15
10
5
0
-5
-10
-15
-20
-25
-30
35
0
5
10
Observation Window (minutes)
Maximum Error
45/48
Trumpet and Flow Rate Curves
15
20
25
Minimum Error
Overall Mean = -0 .3%
15
20
25
Minimum Error
Overall Mean = 0 .0%
30
35
30
35