Pole Clamp Installation; Docking Station/Workstation* Or Equipment Rail Installation - CareFusion Alaris VP Plus Guardrails Directions For Use Manual

Volumetric pump
Table of Contents

Advertisement

Pole Clamp Installation

The pole clamp is fitted to the rear of the Pump and will provide secure fixing to vertical IV poles of a diameter between 15 and 40mm .
1 . Pull the folded pole clamp towards you and unscrew the clamp to
leave enough room for the size of the pole .
2 . Place Pump around pole and tighten screw until the clamp is
secured to the pole .
Never mount the Pump such that the IV infusion stand
A
becomes top heavy or unstable.
Ensure the pole clamp is folded away and stored
within the recessed area at the rear of the Pump
before connecting to a Docking Station/Workstation*
or when not in use.

Docking Station/Workstation* or Equipment Rail Installation

The rotating cam can be fitted to the rectangular bar on the Docking Station/Workstation* or equipment rails measuring 10mm by
25mm .
1 . Align the rotating cam on the rear of the Pump with the
rectangular bar on the Docking Station/Workstation* or
the equipment rail .
2 . Push the Pump firmly onto the rectangular bar or
equipment rail .
3 . The Pump should click into position when fitted to the
bar .
4 . Ensure that the Pump is positioned securely . Verify Pump
is secure by gently pulling the Pump away from the
Docking Station/Workstation* without using the release
lever . When the Pump is securely attached, it should not
come off the Docking Station/Workstation* .
5 . To release, push the release lever and pull the Pump
forwards .
Warning: Pump may fall off the Docking Station/Workstation* if not properly mounted which could result in user and/or
patient harm.
* Alaris DS Docking Station and Alaris Gateway Workstation .
It is recommended that infusion bags be located on a hanger directly above the Pump with which they are being
A
used. This minimises the potential for confusion of infusion sets when multiple volumetric pumps are used.
1000DF00660 Issue 9
Alaris™ VP Plus Guardrails™ Volumetric Pump
Recessed area
*
Rectangular bar
13/48
Getting Started
*
Release lever (push
to release)
Rotating cam

Hide quick links:

Advertisement

Table of Contents
loading