Large Volume Pump Configuration Settings

AIL Limit
Bolus Mode
Bolus Rate Default
Bolus Rate Max
Bolus Volume Max
Infusion Rate Max
KVO Rate
Near End of Infusion Point
Primary VTBI Max
Prime Rate
Prime Volume Max
Secondary Infusion
Sec. Infusion Rate Max
Sec. VTBI Max
Downstream Occlusion Alarm
Pressure
Downstream Pressure Maximum
Auto Pressure
Auto Set Pressure
Auto Offset
Alarm Volume Ramping
Disable Near End Of Infusion
Set the single bubble Air-in-line alarm .
Controls whether or not the Pump allows bolus delivery method .
Set the default value for bolus rates .
Set the maximum permissible bolus rate .
Set the maximum permissible bolus volume .
Set the maximum permissible infusion rate .
Set the maximum infusion rate when performing KVO delivery .
Set the point defined in terms of remaining time at which the infusion is deemed to be Near end
i .e . nearly complete .
Set the maximum VTBI for primary infusions .
Set the rate at which priming is performed .
Set the maximum volume that can be purged as part of a given prime operation .
Allows the use of a secondary infusion in the same channel .
Set the maximum permissible infusion rate for secondary infusions .
Set the maximum permissible setting for the Volume To Be Infused for secondary infusions .
Sets the default downstream occlusion alarm limit .
Sets the maximum permitted downstream occlusion alarm limit .
Controls the setting of the downstream occlusion alarm limit offset above the measured
pressure at that moment .
Controls whether the downstream occlusion alarm limit can be set automatically .
When the downstream occlusion alarm limit is set automatically, this value is the amount added
to the stable running downstream pressure to set the limit .
Controls whether the alarm volume ramping feature is enabled, which controls whether the
alarm, warning or prompt volume increases over time .
Controls whether the Near End of Infusion warning is disabled .