Specifications

Electrical Protection
Class I, Type CF (Defibrillation-proof )
Electrical/Mechanical Safety
Complies with IEC/EN60601-1 and IEC/EN60601-2-24.
Electro Magnetic Compatibility (EMC)
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.
Electrical Safety
Typical earth leakage current 78µA.
Typical Enclosure Leakage Current (Normal Condition) = 0µA
Typical Protective Earth Resistance = 32mOhms
The above measurements are for guidance only, IEC/EN60601-1 limits are defined below:
Earth Leakage Current (Normal Condition) <= 500µA
Enclosure Leakage Current (Normal Condition) <= 100µA
Protective Earth Resistance <= 200mOhms
Classification
Continuous mode of operation, Portable Equipment
AC Power Supply
100 - 230 VAC, 50 - 60Hz, 60VA (Maximum).
Fuse Type
2 X T 1.25 A, slow blowing.
Dimensions
148mm (w) x 225mm (h) x 148mm (d). Weight: approx. 2.5kg (excluding power cable).
Protection against fluid ingress
IPX3 - Protected against direct sprays up to 60° from the vertical.
Environmental Specifications
Condition
Temperature
Humidity
Atmospheric Pressure
*Non condensing.
Battery Specifications
Rechargeable NiMH (Nickel Metal Hydride). Automatically charges when the pump is connected to AC power.
• Battery Life - For a 24 hour battery charge time, the pump at 25ml/h will have a Mean Time To Battery Empty of 6 hours.
• Battery Charging - 2.5 hours to 95%.
Memory Retention
The electronic memory of the pump will be retained for more than 2 years with normal use.
Operating
+5°C - +40°C
20% - 90%*
700hPa - 1060 hPa
1000DF00546 Issue 5
Alaris™ GP (Guardrails™) Volumetric Pump
Transport & Storage
-20°C - +50°C
15% - 95%*
500hPa - 1060hPa
42/50
Specifications