Table 6.8 Continuous Mode Parameters; Table 6.9 Continuous Mode Operational Sequence - Emerson Branson LPX Instruction Manual

Table of Contents

Advertisement

6.5.3
Continuous Mode
In this mode, ultrasonics are applied to the sample until you stop the cycle. The user
controls the timing of ultrasonics manually. The power supply starts when a START
condition is received and stops when a STOP condition is received.
The following table shows the parameters of Continuous mode, along the default, max.
and min. values.
Table 6.8
Continuous Mode Parameters
Parameter
Amplitude
Ground Detect
Entering an illegal value will generate 3 beeps. The system will not accept out-of-range
parameters. (See
NOTICE
You can return to the ready screen without saving any change by pressing the ESC key
at any time.
Table 6.9
Continuous Mode Operational Sequence
Step
Turn on power, wait for the LCD to show ready screen and
1
current mode.
Press the Mode/Configuration key once to get into mode
2
selection. Current selected mode icon will blink.
100-412-196 REV. 06
Default
50%
0.05 s
7.4 Alarms/Errors
for further details).
Action
Max. Value
100%
10%
0.99 s
0.00 s
Min. Value
Reference
61