5.3

Assembling the Equipment

The LPX Power Supply unit is pre-assembled and requires no special tools, however other
components must be connected to the unit in order for the system to operate. Some
assembly of the ultrasonic horn is required, as described in the following sections.
5.3.1
Setup Procedure
To set up your LPX Power Supply, take the following steps:
Table 5.9
Setup Procedure
Step
1
2
3
4
To remove a horn, use spanner wrenches supplied in the appropriate kit. Never attempt to
remove a horn by holding the converter housing in a vise. If necessary, secure the largest
portion of the horn in a soft-jawed vise. See
Converters.
38
Connect the tip, horn, and converter, following the procedure in
Connecting Tips, Horns, and
Mount the converter/horn assembly in a stand, into the proper hand held
device, or other suitable support.
Set the ON/OFF switch on the back of the unit to the OFF position.
Plug the line cord into the unit, and then into an appropriate electrical
power outlet, ensuring that the power supply is grounded to prevent
electrical shock.
NOTICE
A plug meeting local electrical codes may be required.
Action
Converters.
5.3.2 Connecting Tips, Horns, and
5.3.2
100-412-196 REV. 06