6.2 Adding a Log (cont.)
Program the new log parameters as shown in Table 13 below:
Parameter
LOG NAME:
FORMAT:
CHANNEL
INTERVAL:
START DATE & TIME:
END DATE & TIME:
Note the following:
The PT900 transmitter will only start the log if the programmed START
DATE & TIME
The programmed END DATE & TIME must occur after the programmed
START DATE & TIME
There is no defined limit for either the total number of logs or for the size
of an individual log stored in the transmitter, but the total available amount
of log storage memory is limited to the PT900 storage capacity.
®
TransPort
PT900 User's Manual
Table 13: Programming the Log Parameters
Entry Type
Manual entry
Slide switch
Drop-down list
Drop-down list
Manual entry
Manual entry
arrives and the transmitter is powered On at that time.
, or an error message is displayed.
Description
11 characters maximum
Linear:
Record all values from start time
to end time
Circular:
After 100 values are recorded,
the oldest value is deleted before a new
value can be added.
CH1
, CH2 , Average , General (see
Table 12 on page 93 for the available
measurement variables).
1-20 sec
(the time between each pair of
consecutive values).
When the log starts
When the log stops
Chapter 6. Logging Data
103