Cleaning/Disinfection procedure

The procedure cleans or disinfects the instrument according to a set of actions. Refer to Chapter 11:
"Maintenance", particularly "Decontamination procedure".
1. Press the OK button on the Cleaning/Disinfection procedure row.
2. Empty all bottles and fill the buffer bottle with cleaning reagent.
Press the OK button.
3. Select the soak time using the Right arrow key and number keys.
The default is 15 min. Press the OK button.
Priming, preparing the soak and soaking take place.
4. Press the F2 key to abort the soaking if you want a shorter soak time
than the set soak time and to continue the procedure. Priming is
then carried out.