Running Calibration on the M9393A from the 89600 VSA Software
It is possible to run calibrations on the M9393A VSA from the 89600 VSA software
when the two are correctly configured. The calibration path in the 89600 VSA
software is
Procedure for a Two Channel Configuration
To create a new M9393A instrument, use the Keysight 89600 VSA Software and
follow these steps:
Keysight M9393A PXIe Performance Vector Signal Analyzer
For additional information on each extension item, highlight the item and a
summary is displayed near the bottom of the extension dialog box.
Utilities > Calibration
.
Before running a calibration on the M9393A, it is important to make
sure that there is no RF signal connected to the RF In connector of
the M9365A Downconverter module in the M9393A instrument. A
signal present at the RF Input connector of the M9365A
Downconverter can result in an unsuccessful calibration of the
M9393A instrument.
For extended frequency range configuration without M9169E,
remember to connect cable between Aux 1 Out and Aux 2 In
connectors before running calibration.