Configurations
WetCalFunction(x)=
Wet calibration
Configure the date points listed below from the Daniel MeterLink Calibration|Meter
Factors menu:
Table 5-2:
Daniel MeterLink
Display Name
Flow calibration method
Procedure
1.
2.
40
selected wet-calibration function
Data Points for wet calibration
Data points, options and guidelines
Data points affected:
CalMethod
d Options:
Piece-wise (2)
Polynomial (1)
None
(0) Guidelines:
This value should only be modified under direction of Daniel per-
sonnel.
Choose a Flow Calibration Method:
Polynomial
Piece-wise linear
None (If set to None then the wet calibration has the same value as the dry
calibration.)
Select the Calibration|Meter Factors menu.
3410 Series Gas Ultrasonic Flow meters