Chapter 8 - Automatic Measurement

UTD2000CEX-II series digital fluorescence oscilloscope can automatically
measure up to 34 parameters. Press the MEASURE button to enter the
automatic measurement menu.
Automatic Measurement menu
Options
Descriptions
Functions
Off
Close all Parameters
All Parameters
Pop-up a frame with all
On
parameters presenting
Open/Close menu for user defined
parameters. When it is on, define
User Defined
On/Off
all required parameters with using
the Multipurpose knob.
Off
Real-time dynamically indicates
Indicator
measurement parameters
Parameter 1~5
Advanced
Enter advanced measurement menu
measurement
Off
Turn off the feature
Automatic calculation of user defined
Measurement
parameters of average, max, and min.
statistics
On
Only applicable when user defined
parameters are present.
Reset statistics
Restart statistics
Clear
Clear all measurement
Advanced measurement menu
Options
Descriptions
Functions
Delay measurement
Measurement
parameters
Phase measurement
Source
FRR: time between the first rising edge of
source 1 and the first rising edge of source 2
FRF: time between the first rising edge of
source 1 and the first falling edge of source 2
FFR: time between the first falling edge of
source 1 and the first rising edge of source 2
FFF: time between the first falling edge of
source 1 and the first falling edge of source 2
edge
LRF: time between the last rising edge of
source 1 and the last edge of source 2
LRR: time between the last rising edge of
source 1 and the last rising edge of source 2
LFR: time between the last falling edge of
source 1 and the last rising edge of source 2
LFF: time between the last falling edge of
source 1 and the last falling edge of source 2
UTD2000CEX-II
User Manual