Device profile
January 2019
Configuration
Concentration measurement
B.3
I/O data
B.3.1
Input assembly for standard configuration
Byte
Access
Name
0–3
Get
Mass flow rate
4–7
Get
Density
8–11
Get
Temperature
12–15
Get
Volume flow rate (liquid
volume)
16–19
Get
Mass total
20–23
Get
Volume total (liquid
volume)
24–27
Get
Drive gain
28–29
Get
Status word
30–31
Get
Status word
32–35
Get
Gas standard volume flow
rate
36–39
Get
Gas standard volume total
B.3.2
Output assembly for standard configuration
Byte
Access
Name
0–3
Get/Set
External Temperature
44
Output parameters (explicit data)
See
Output parameters (explicit data) for concentration
measurement configuration
Type
Notes
Float
Float
Float
Float
Valid only when Gas Standard Volume is not
enabled.
Float
Float
Float
U16 or Word
•
For Model 1700 Analog, Model 2700 Analog,
Model 1500 Analog, Model 2500, and all Series
3000 transmitters: SNS Status Word 1 (see
Status Word
•
For MVD Direct Connect and 9739 MVD
transmitters: SNS Status Word 2 (see
Word
U16 or Word
•
For Model 1700 Analog, Model 2700 Analog,
Model 1500 Analog, Model 2500, and all Series
3000 transmitters: SNS Status Word 2 (see
Status Word
•
For MVD Direct Connect and 9739 MVD
transmitters: SNS Status Word 1 (see
Word
Float
Valid only when Gas Standard Volume is enabled.
Float
Valid only when Gas Standard Volume is enabled.
Type
Notes
Float
MMI-20019808
1)
2)
2)
1)
Micro Motion EtherNet/IP Module
User Guide
Status