Thies CLIMA DLx-MET Instructions For Use Manual page 21

Table of Contents

Advertisement

4. SENSOR 1 Wind speed:
* Windspeed:
NN.N m/s
Classic
I:N
Line 2:
Line 3:
Serial commands: "WV" and "WW" (ref. to chap.5.5)
Measuring range:
Resolution:
5. SENSOR 2 Wind direction:
* Winddirection:
NNN °
5-/8-Bit
Line 2:
Line 3:
Serial command: "WD" (ref. to chap.5.5)
Measuring range:
Resolution:
6. SENSOR 3 Temperature 1:
Temperature 1:
NNN.N °C
Display of the instantaneously measured temperature1.
Measuring range:
Resolution:
Display of the instantaneously measured wind speed.
Display and input type of sensor (Compact [4.3519.X], Classic [4.3303.X],
First Class [4.3350.X] and current flow test of sensor (I:0 -> off I:1 -> on)
0.5 ... 65.0 m/s (Compact)
0.3 ... 50.0 m/s (Classic)
0.3 ... 75.0 m/s (First Class)
0.1 m/s
Display of the instantaneously measured wind direction.
Display and input type of sensor (5-/8-Bit syn.ser. or 10-Bit syn.ser.)
5-Bit sensor:
0 ... 349°
<= 12°
-40 ... 70 °C
0.1 °C
8-Bit sensor:
0 ... 357°
<= 3°
21 - 56
10-Bit sensor:
0 ... 359°
021579/06/09