Sartorius PMA.World Installation Instructions Manual page 40

Electronic paint-mixing scales for use in zone 1 hazardous areas
Table of Contents

Advertisement

Non-hazardous area
Hazardous area
Zone 1
Gas: IIB T4
Remarks:
1: Combined circuits
2: Measured against ground
3: Between signal lines
4: Only with the PMA7500.-X..G.
*: Ohmically limited
Date
Name
Erstellt
10.01.06
Klausgrete
Written by
Geprüft
10.01.06
Klausgrete
Reviewed by
Freigabe
10.01.06
Klausgrete
Released by
40
YCO11-Z
EX-link box
II (2)G [EEx ib] IIB
Sartorius cable; permanently
mounted on the Ex-link box
(max. length: 100m)
Differential data
transmission is
only possible
with the PMA
7500.X.GL (in
power supply
cable)
9-contact male
4
connector
PMA7501.-X....
II 2 G EEx ib IIB T4
Material
PMA7501.-X.... /
Differential Data Transfer
Ausgabe / Revision
Änderung / Alteration
00
---
Ethernet or RS-232 data
interface
for connecting to
equipment using 250V
power supply, maximum
TM01-X
II 2 G EEx ib IIB T4
9-contact male
connector
RS-232 data output port
Vi
12.6 V
Ii
132 mA*
Pi
412 mW
Ci 2 nF
Li
0
RS-232 data output port
2
Vi
12.6 V
3
25.2 V
Ii
330 mA*
Pi
any
Ci 1 nF
Li
0
9-contact
female
connector
Bennenung / Title
Verification of Intrinsic Safety
Zeichnungs-Nr. / Drawing number
35958-741-60-A4
1
Vo
12.0 V
Io
128 mA
Po
376 mW
Co
9 µF
Lo
50 mH
1
2
Vo
12.0 V
3
24.0 V
Io
125 mA
Po
373 mW
2
Co
9 µF
3
0.93 µF
Lo
8 mH
Maßstab /
Scale
---
Blatt
3
Sheet
von
3
of

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pma.net