Sartorius PMA.World Installation Instructions Manual page 11

Electronic paint-mixing scales for use in zone 1 hazardous areas
Table of Contents

Advertisement

Weighing with Two Decimal Places
Note:
To weigh using two decimal places, you must first
adapt the settings (refer to the chapter entitled "Menu
Settings")
Press the I key (6). The display shows "0.00 g."
Place an empty paint can on the weighing pan (11).
Press the U key (5). The display shows "0.00 g."
Pour in the first component: 205.50 g.
Read off the weight as soon as the stability symbol
appears; in this case, "g."
Pour in additional components until the desired weight
of your formula is reached.
Remove the filled paint can from the weighing pan.
Important Note:
If you tare the scale, and then press the I key (6) to
toggle to the second decimal place with a resolution of
0.05 g, you can continue weighing with two decimal
places up to 999.95 g.
For weights exceeding 999.95 g, only one decimal place
will be displayed.
Never use a hammer to close the lid of a paint can
while it is still on the weighing pan.
Otherwise, you will damage the weighing system
11